Search In this Thesis
   Search In this Thesis  
العنوان
The effect of manufacturing technique, storage and sterilization on the accuracy of computer aided implant surgical guides :
الناشر
Mahetab Hamdi Abdelfatah ,
المؤلف
Mahetab Hamdi Abdelfatah
هيئة الاعداد
باحث / Mahetab Hamdi Abdelfatah
مشرف / Ashraf Abd El Monem Amer
مشرف / Amr Hosny Elkhadem
مشرف / Aml fathi kadah
مناقش / Iman El Taftazani
تاريخ النشر
2016
عدد الصفحات
109 :
اللغة
الإنجليزية
الدرجة
ماجستير
التخصص
Dentistry (miscellaneous)
تاريخ الإجازة
11/10/2016
مكان الإجازة
جامعة القاهرة - الفم والأسنان - Prosthodontics
الفهرس
Only 14 pages are availabe for public view

from 129

from 129

Abstract

This invitro study was designed to compare the effect of four manufacturing techniques, storage, chemical sterilization and autoclaving on accuracy of computer guided surgical guides. An impression for class III Kennedy classification partially edentulous case was made and poured into epoxy resin to construct a cast. The cast was laser scanned and a surgical guide was designed by the Blue sky software for implant planning. Four techniques were used for manufacturing which were SLS, milling, SLA and FDM. Measuring accuracy were done in three terms; evaluation of sleeve hole tolerance, guide adaptation and accuracy of sleeve angle. The sleeve hole tolerance was evaluated by special customized metal cylinders of different diameters where each diameter was used to evaluate its fit inside the sleeve hole. For the microgap photos were captured by digital microscope and then analyzed by image J software. Regarding angular deviation, laser scanning of the surgical guide with special customized metal cylinders both were sealed on the model was done then evaluating of angular deviation was carried by blue sky..The accuracy was measured in five phases after 24 hour of manufacturing, one week storage, two week storage, chemical sterilization and autoclaving. All results were tabulated and analyzed to reach conclusion